Minimally invasive glaucoma surgery

Micro-invasive glaucoma surgery (MIGS) is the latest advance in surgical treatment for glaucoma, which aims to reduce intraocular pressure by either increasing outflow of aqueous humor or reducing its production. MIGS comprises a group of surgical procedures which share common features.[1] MIGS procedures involve a minimally invasive approach, often with small cuts or micro-incisions through the cornea that causes the least amount of trauma to surrounding scleral and conjunctival tissues. The techniques minimize tissue scarring, allowing for the possibility of traditional glaucoma procedures such as trabeculectomy or glaucoma valve implantation (also known as glaucoma drainage device) to be performed in the future if needed.[2]

Traditional glaucoma surgery generally involves an external (ab externo) approach through the conjunctiva and sclera; however, MIGS procedures reach their surgical target from an internal (ab interno) route, typically through a self-sealing corneal incision. By performing the procedure from an internal approach, MIGS procedures often reduce discomfort and lead to more rapid recovery periods.[1][2] While MIGS procedures offer fewer side effects, the procedures tend to result in less intraocular pressure (IOP) lowering than with trabeculectomy or glaucoma tube shunt implantation.[3]

Medical uses

Glaucoma is a group of eye disorders in which there is a chronic and progressive damage of the optic nerve.[4] Increased intraocular pressure (IOP) is the main and only modifiable risk factor, attributed to the progression of the disease. During the last 25 years, glaucoma management has been based in the used of pharmaceutical therapies and incisional surgery.[5] MIGS procedures can provide the patient sustained IOP reduction while minimizing the risk and complications associated with glaucoma interventions and decrease the dependence of glaucoma medications.[1]

Adverse events

MIGS procedures offer an excellent safety profile, with minimal incidence of complications, especially when compared with other forms of glaucoma surgery.[1][2][6]

Procedures

MIGS objective, like all glaucoma surgeries, is to achieve lowering of IOP by either increasing aqueous humour outflow, the fluid that is produced by the eye and fills the space between the cornea and the lens, or decreasing the production of aqueous humour. MIGS can be divided in procedures that lower IOP by using micro-stents and minimally invasive procedures.

Micro-stents/Shunt Devices

iStent (Glaukos Corp., Laguna Hills, CA)

The iStent Trabecular Micro-Bypass Stent, or simply iStent, is the smallest implantable medical device, designed to lower intraocular pressure by facilitating trabecular outflow of aqueous fluid.[7] The trabecular outflow is one of the major outflow pathways for aqueous humor in the eye and has been the target of both pharmaceutical and surgical therapeutic approaches in glaucoma.

The 1-millimeter long iStent is a titanium device inserted via an internal approach through the trabecular meshwork into Schlemm’s Canal, bypassing the trabecular meshwork and facilitating flow of aqueous from the eye.[1][2][3] Studies have shown that the iStent is an effective procedure, typically lowering intraocular pressure to the mid-teens.[8][9]

The iStent is the first MIGS device to get FDA approval for implantation in combination with cataract surgery.[10] The device has also been shown to offer better IOP control than cataract surgery alone up to one year of follow-up in a large randomized controlled FDA study, although the effectiveness was significantly reduced by 2 years.[11][12] Safety of the iStent was comparable to cataract surgery alone which is much better than conventional trabeculectomy.[11][12][13][14] Common complications include failure to implant the device, touching the iris with the device, and touching the undersurface of the cornea (endothelium) with the device.[15] Multiple studies have since confirmed the MIGS-type efficacy and safety profile of the iStent.[15][16]

To address the reduced effectiveness at 2 years, some studies have been performed with multiple iStents.[17]

CyPass Micro-Stent (Transcend Medical, Inc., Menlo Park, CA)

The CyPass Micro-Stent is the first MIGS device developed for lowering of IOP through the suprachoroidal space (virtual space between the choroid and sclera created by implantation of the device), a part of the uveoscleral outflow pathway for aqueous humor. The uveoscleral pathway is an important pathway for aqueous outflow and drainage which can account for up to 57% of total outflow.[18] Cyclodialysis cleft procedures were initially used to access this pathway with significant IOP lowering, but the cleft was prone to high anatomic variability as well as early postoperative closure due to the lack of a permanent drainage implant with a standardized and uniform conduit.[19]

It has a microlumen of 300 microns and is designed to augment outflow to the suprachoroidal space in order to control intraocular pressure; it is indicated for the treatment of primary open-angle glaucoma.[20] The stent is implanted through an ab interno approach and inserted into the supraciliary space (between the ciliary body and sclera), effectively creating a controlled cyclodialysis cleft, which is kept open by the device.[21][22]

The first clinical data on the device was presented in 2010 showed sustained IOP lowering and MIGS-like safety profile.[23] This has been substantiated in subsequent studies in the combined setting with cataract surgery and as a stand-alone treatment for patients failing glaucoma topical therapy.[6][21][22][24] Data from a large randomized controlled pivotal study (The COMPASS Clinical Study) has reported positive efficacy after 2 years of follow-up and will be submitted to FDA for approval.[25][26] The CyPass device has been CE-marked since 2009.

Hydrus (Ivantis, Inc., Irvine, CA)

The Hydrus Microstent is an implantable MIGS device for the treatment of primary open angle glaucoma; implantation of this device can be performed in conjunction with cataract surgery.[27] The Hydrus Microstent is the longest of the MIGS devices (8-millimeter long implant), and similar to the iStent it is designed to increase trabecular outflow.

The implant is inserted through the trabecular meshwork, thus bypassing resistance of aqueous flow through the tissue. However, other glaucoma surgeries, such as canaloplasty, have shown that mechanical dilation of Schlemm’s Canal is also associated with a reduction in intraocular pressure.[28] The Hydrus Microstent takes advantage of this property by mechanically dilating and holding open Schlemm’s Canal. The length of the Hydrus Microstent is thought to open approximately one quarter of Schlemm’s Canal, routing aqueous into open downstream collector channels.

Clinical data from a randomized controlled study demonstrates efficacy of the device at 1 and 2 years.[29] The Hydrus Microstent is currently being investigated in an FDA approved study for mild-to-moderate glaucoma.

XEN Gel Stent (AqueSys, Inc., Aliso Viejo, CA)

The XEN Gel Stent is an implantable transcleral microsurgical device that allows the aqueous fluid to drain from the anterior chamber into the subconjunctival space, a pathway utilized by traditional trabeculectomy and glaucoma drainage device surgeries.[30] Unlike the latter two procedures, the XEN Gel Stent is performed through an internal approach and avoids directly incising and disrupting the conjunctiva itself.[31][32]

The 6-millimeter stent is placed through the trabecular meshwork, with one end of the stent sitting directly underneath the conjunctiva, past the outer wall of the sclera. The inner tip of the stent sits in the anterior chamber, allowing a direct connection for aqueous to flow into the subconjunctival space. The stent is made of a non-bioreactive material, which does not promote scarring of the conjunctiva.[31] The XEN Gel Stent is currently under FDA investigation for refractory glaucoma and has not yet been approved for clinical use in the United States.

There are currently no published peer-reviewed clinical studies reporting results from the XEN device.[33]

Minimally Invasive Procedures

Endocyclophotocoagulation

Aqueous humor is produced in the portion of the eye known as the ciliary body. The ciliary body contains 72 protrusions known as ciliary processes, from which the aqueous is produced. The destruction of these ciliary processes with a diode laser, known as cyclophotocoagulation, can be used to decrease the amount of aqueous humor produced, thereby reducing the intraocular pressure.

Cyclophotocoagulation traditionally has been performed using an external laser through the sclera, known as transcleral cyclophotocoagulation. However, side effects of the transcleral approach can include significant inflammation, chronically low intraocular pressure, intraocular bleeding, and permanent shutdown of the eye, known as phthisis.[34][35] Recent advances have now allowed a diode laser to be combined with a camera (endoscope) allowing for direct visualization of the ciliary processes during the ablation (Endo Optiks, Beaver Visitec, Waltham, MA).[36]

Endocyclophotocoagulation is indicated for the treatment of both open and closed angle glaucoma and is performed in eyes which have already undergone cataract surgery or performed concomitantly with cataract removal. The largest investigation of endocyclophotocoaguation has shown a significant decrease in intraocular pressure of up to 10 mmHg, as well as a significant reduction in number of glaucoma medications needed. Reported adverse reactions include intraocular inflammation, bleeding, and cystoid macular edema (swelling of the retina).[37]

Trabectome (NeoMedix Corp., Tustin, CA)

The Trabectome, or trabeculectomy ab interno, is a microsurgical device cleared by the U.S. Food and Drug Administration since 2006, used in patients with open angle glaucoma to excise a strip of trabecular meshwork, the tissue primarily responsible for the increased resistance of aqueous outflow in glaucoma.[38] The Trabectome uses electrocautery via an internal approach to vaporize the trabecular meshwork, creating a large pathway for aqueous to flow, with minimal trauma to surrounding tissues. The procedure can be performed alone or in conjunction with cataract surgery.[39]

Trabectome is unique among the MIGS procedures, as there is no physical device implanted inside the eye; the pressure lowering is a direct result from the destruction and removal of the trabecular meshwork. Studies have found a decrease in intraocular pressure to the mid-teens following the procedure, which has a favorable safety profile.[39][40] The most common complication from Trabectome surgery is bleeding, which can blur vision and take extra time to recover. The surgery site can scar over time and the pressure can go back up.[14]

In early 2014, the NeoMedix received a warning letter from FDA regarding marketing practices.[41]

References

  1. 1 2 3 4 5 Saheb H, Ahmed II. Micro-invasive glaucoma surgery: current perspectives and future directions. Curr Opin Ophthalmol. 2012 Mar;23(2):96-104. doi: 10.1097/ICU.0b013e32834ff1e7.
  2. 1 2 3 4 Minimally Invasive Glaucoma Surgeries (MIGS)
  3. 1 2 "Update on Minimally Invasive Glaucoma Surgery (MIGS) and New Implants".
  4. Casson, Robert (5 April 2012). "Definition of glaucoma: clinical and experimental concepts". Clinical & Experimental Ophthalmology. doi:10.1111/j.1442-9071.2012.02773.x.
  5. Jampel, Henry D. (22 June 2015). "A Quarter Century's Progress in the Treatment of Open-Angle Glaucoma". Ophthalmology. doi:10.1016/j.ophtha.2014.10.005.
  6. 1 2 Hoh, H.; Grisanti, S.; Rau, M.; Ianchulev, S. (25 April 2014). "Two-year Clinical Experience with the CyPass Micro-Stent: Safety and Surgical Outcomes of a Novel Supraciliary Micro-Stent". Klin Monatsbl Augenheilkd. doi:10.1055/s-0034-1368214.
  7. "Glaukos: iStent".
  8. Villalobos-Arriola, P; Martinez-de-la-Casa, JM; Garcia-Feijoo, J. (20 April 2013). "Mid-term evaluation of the new Glaukos iStent with phacoemulsification in coexistent open-angle glaucoma or ocular hypertension and cataract.". British Journal of Ophthalmology. doi:10.1136/bjophthalmol-2012-302394. PMID 23603758.
  9. Fea, Antonio M. (November 2008). "Phacoemulsification versus phacoemulsification with micro-bypass stent implantation in primary open-angle glaucoma". Journal of Cataract & Refractive Surgery. doi:10.1016/j.jcrs.2009.10.031.
  10. "FDA: Glaukos iStent".
  11. 1 2 Samuelson, Thomas W.; Katz, Jay; Wells, Jeffrey M.; Duh, Yi-Jing (September 15, 2010). "Randomized Evaluation of the Trabecular Micro-Bypass Stent with Phacoemulsification in Patients with Glaucoma and Cataract". Ophthalmology.
  12. 1 2 Craven, Randy (March 2011). "Cataract surgery with trabecular micro-bypass stent implantation in patients with mild-to-moderate open-angle glaucoma and cataract: Two-year follow-up". Journal of Cataract & Refractive Surgery.
  13. "Surgical Complications in the Tube Versus Trabeculectomy Study During the First Year of Follow-up - American Journal of Ophthalmology". www.ajo.com. Retrieved 2016-01-06.
  14. 1 2 "Postoperative Complications in the Tube Versus Trabeculectomy (TVT) Study During Five Years of Follow-up - American Journal of Ophthalmology". www.ajo.com. Retrieved 2016-01-06.
  15. 1 2 "Cataract surgery with trabecular micro-bypass stent implantation in patients with mild-to-moderate open-angle glaucoma and cataract: Two-year follow-up - Journal of Cataract & Refractive Surgery". www.jcrsjournal.org. Retrieved 2016-01-06.
  16. Wellik, Sarah; Dale, Elizabeth. "A review of the iStent® trabecular micro-bypass stent: safety and efficacy". Clinical Ophthalmology. doi:10.2147/opth.s57217.
  17. Belovay, Graham W.; Naqi, Abdulla; Chan, Brian J.; Rateb, Mahmoud; Ahmed, Iqbal Ike K. "Using multiple trabecular micro-bypass stents in cataract patients to treat open-angle glaucoma". Journal of Cataract & Refractive Surgery. 38 (11): 1911–1917. doi:10.1016/j.jcrs.2012.07.017.
  18. Toris, CB (2008). "Chapter 8: Aqueous Humor Dynamics II – Clinical Studies". Current Topics in Membranes.
  19. Heine L. Die Cyclodialyse, eine neue glaukomoperation. Dtsch Med Wschr. 1905:824-826
  20. Ianchulev, Tsontcho (2014-01-01). MD, John R. Samples; FRCSC, Iqbal Ike K. Ahmed MD, eds. Suprachoroidal Space as a Therapeutic Target. Springer New York. pp. 33–43. doi:10.1007/978-1-4614-8348-9_3. ISBN 978-1-4614-8347-2.
  21. 1 2 Garcia-Feijoo, J., Rau, M., Ahmed, I., Antonio, A., Grabner, G., & Ianchulev, T. (2012, September). Safety and efficacy of CyPass Micro-Stent as a stand-alone treatment for open-angle glaucoma: worldwide clinical experience. In European Society of Cataract and Refractive Surgeons annual meeting, Milan (pp. 8-12)
  22. 1 2 Saheb, Hady; Ianchulev, Tsontcho; Ahmed, Iqbal ‘Ike’ K. (2014-01-01). "Optical coherence tomography of the suprachoroid after CyPass Micro-Stent implantation for the treatment of open-angle glaucoma". British Journal of Ophthalmology. 98 (1): 19–23. doi:10.1136/bjophthalmol-2012-302951. ISSN 1468-2079. PMID 23743436.
  23. "Minimally Invasive Ab-Interno Suprachoroidal Device (CyPass) for IOP Control in Open-angle Glaucoma - Transcend Medical". Transcend Medical. Retrieved 2016-01-06.
  24. Grisanti, PD Dr S.; Margolina, E.; Hoeh, H.; Rau, M.; Erb, C.; Kersten-Gomez, I.; Dick, H. B.; Grisanti, S. (2013-08-21). "Supraziliarer Mikrostent bei Offenwinkelglaukom". Der Ophthalmologe (in German). 111 (6): 548–552. doi:10.1007/s00347-013-2927-6. ISSN 0941-293X.
  25. "Transcend Medical touts data, plans FDA filing for CyPass glaucoma stent – MassDevice". MassDevice. Retrieved 2016-01-06.
  26. "Startup Transcend to submit PMA in second half for glaucoma micro-stent after positive pivotal data". FierceMedicalDevices. Retrieved 2016-01-06.
  27. "Hydrus™ Procedure | Ivantis". www.ivantisinc.com. Retrieved 2016-01-06.
  28. Hays, Cassandra L.; Gulati, Vikas; Fan, Shan; Samuelson, Thomas W.; Ahmed, Iqbal Ike K.; Toris, Carol B. "Improvement in Outflow Facility by Two Novel Microinvasive Glaucoma Surgery Implants". Investigative Ophthalmology & Visual Science. 55 (3). doi:10.1167/iovs.13-13353. PMC 3973188Freely accessible. PMID 24550367.
  29. Pfeiffer, Norbert; Garcia-Feijoo, Julian; Martinez-de-la-Casa, Jose M.; Larrosa, Jose M.; Fea, Antonio; Lemij, Hans; Gandolfi, Stefano; Schwenn, Oliver; Lorenz, Katrin. "A Randomized Trial of a Schlemm's Canal Microstent with Phacoemulsification for Reducing Intraocular Pressure in Open-Angle Glaucoma". Ophthalmology. 122 (7): 1283–1293. doi:10.1016/j.ophtha.2015.03.031.
  30. "XEN Gel Stent from AqueSys, Inc.". www.aquesys.com. Retrieved 2016-01-06.
  31. 1 2 Lewis, Richard A. "Ab interno approach to the subconjunctival space using a collagen glaucoma stent". Journal of Cataract & Refractive Surgery. 40 (8): 1301–1306. doi:10.1016/j.jcrs.2014.01.032.
  32. . Vera, V. I., & Horvath, C. (2014). XEN Gel Stent: The Solution Designed by AqueSys®. In Surgical Innovations in Glaucoma (pp. 189-198). Springer New York
  33. "FDA" (PDF).
  34. Allingham RR, Damji K, Freedman S, Moroi S, Shafranov G. Shields Textbook of Glaucoma. 5th ed. Baltimore, MD: Lippincott, Williams and Wilkins; 2005. Chapter 43. Cyclodestructive Surgery; pp. 644–61.
  35. Kahook MY, Noecker RJ, Schuman JS. Albert and Jakobiec's Principles and Practice of Ophthalmology. 3rd ed. Saunders Elsevier; 2008. Cyclophotocoagulation; pp. 2871–4.
  36. Siegfried, CarlaJ; Falkenberry, SuzanneM (2009-07-01). "Endocyclophotocoagulation". Middle East African Journal of Ophthalmology. 16 (3). doi:10.4103/0974-9233.56225. PMC 2813596Freely accessible. PMID 20142978.
  37. Chen, J.; Cohn, R. A.; Lin, S. C.; Cortes, A. E.; Alvarado, J. A. (1997-12-01). "Endoscopic photocoagulation of the ciliary body for treatment of refractory glaucomas". American Journal of Ophthalmology. 124 (6): 787–796. doi:10.1016/s0002-9394(14)71696-4. ISSN 0002-9394. PMID 9402825.
  38. Francis, BA, Singh K, Lin SC, Hodapp E, Jampel H, Samples J, Smith SD. Novel Glaucoma Procedures. Ophthalmology [0161-6420] Francis yr:2011 vol:118 iss:7 pg:1466 -1480.
  39. 1 2 Minckler, Don; Mosaed, Sameh; Dustin, Laurie; Ms, Brian Francis (2008-01-01). "Trabectome (trabeculectomy-internal approach): additional experience and extended follow-up". Transactions of the American Ophthalmological Society. 106: 149–159; discussion 159–160. ISSN 1545-6110. PMC 2646453Freely accessible. PMID 19277230.
  40. Jordan, Jens F.; Wecker, Thomas; Oterendorp, Christian van; Anton, Alexandra; Reinhard, Thomas; Boehringer, Daniel; Neuburger, Matthias (2013-10-26). "Trabectome surgery for primary and secondary open angle glaucomas". Graefe's Archive for Clinical and Experimental Ophthalmology. 251 (12): 2753–2760. doi:10.1007/s00417-013-2500-7. ISSN 0721-832X. PMC 3889259Freely accessible. PMID 24158374.
  41. "FDA Warning Letter Neomedix".
This article is issued from Wikipedia - version of the 11/8/2016. The text is available under the Creative Commons Attribution/Share Alike but additional terms may apply for the media files.